ALARMS, SAFETY, AND SELF-DIAGNOSIS
Table 3 Alarms & safety function list
Alarm & safety
Situation
If the thermal sensor is
disconnected.
If the thermal sensor is
short-circuited.
If the condenser sensor A
is disconnected.
Sensor abnormality
If the condenser sensor A
is short circuited.
If the condenser sensor B
is disconnected.
If the condenser sensor B
is short-circuited.
If the ambient tem. sensor
is disconnected.
If the ambient tem. sensor
is short-circuited.
When the battery switch
Battery switch check
for power failure alarm is
OFF.
When the fan motor for
Condenser A temp.
cooling the compressor
abnormality
fails.
When the ambient
temperature exceeds the
Condenser B temp.
usable environmental
abnormality
condition, etc.
When communication
between LCD touch
Communication
panel and control
error
substrate is died out or
unstable.
The performance of the
cooling circuit "A" gets
Cooling circuit
abnormality
worse.
(Self-diagnosis time:
The performance of the
2 a.m.~6 a.m.)
cooling circuit "B" gets
worse.
●When "disconnecting/short-circuit of the thermal sensor" and "condenser temp. abnormality" are
activated at the same time, safety operation of compressor OFF is prioritized.
(Ex.) When "disconnecting of the thermal sensor" and "condenser A temp. abnormality" are activated at
the same time;
Compressor A: OFF
Compressor B: Continuous running
●The self-diagnosis time of "cooling circuit abnormality" cannot be changed. Depending on the operating
environment and condition, self-diagnosis time may be shifted or this function may not perform.
48
LCD touch panel
Message display field
Err01:
Temperature Sensor Open.
Err02:
Temperature Sensor Short.
Err03:
Condenser Sensor 'A' Open.
Err04:
Condenser Sensor 'A' Short.
Err05:
Condenser Sensor 'B' Open.
Err06:
Condenser Sensor 'B' Short.
Err07:
Ambient Temp Sensor Open.
Err08:
Ambient Temp Sensor Short.
Err09:
Battery Switch is OFF.
Err20:
Condenser 'A' Temp Abnormal.
Err21:
Condenser 'B' Temp Abnormal.
Err56:
Communication Failure.
Warning:
Cooling Circuit 'A' Abnormal
Warning:
Cooling Circuit 'B' Abnormal
Buzzer
Alarm display
"Warning" is
displayed
alternately in
normal
Intermittent
characters and
tone
reverse video.
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
‫״‬
-----
-----
"Warning" is
displayed
alternately in
Intermittent
normal
tone
characters and
reverse video
-----
-----
"Warning" is
displayed
alternately in
Intermittent
normal
tone
characters and
reverse video.
Remote
Safety
alarm
operation
Unit is
ON
continuous
running.
‫״‬
-----
‫״‬
-----
‫״‬
-----
-----
-----
Compressor
A OFF
ON
Compressor
B OFF
-----
-----
ON
-----