6. Using the ventilator
6.1 Switching on / self test
22
EN
Using the ventilator
1. Open the valve of the oxygen cylinder
The pressure gauge will now show the pressure
in the cylinder.
2. Where appropriate, calculate how long the re-
maining oxygen will last (see "6.11 Calculation
of oxygen content/remaining operating time"
on page 37). Always change the cylinder in good
time, e.g. when the pressure is lower than
50 bar, to ensure that oxygen is available for an
adequate period.
3. Select the desired ventilation settings (see "6.2
Selecting the ventilation settings" on page 23).
4. Switch on the MEDUMAT Easy with ON/OFF
switch. The ventilator will then run a self test
lasting approx. 2 seconds.
If audio response is enabled, you will hear before
the self test the sentence "Open oxygen cylin-
der".
.
slowly