10. Servicing
Interval
Every 2 years (service
and safety check)
Every 4 years
Every 10 years
56
EN
Servicing
Have the cleaned and disinfected device serviced at
regular intervals. Servicing, safety checks ([sicher-
heitstechnische Kontrollen or STKs] in accordance
with §6 of the German law governing the owners/
operators of medical devices - only applies to Germa-
ny) and maintenance measures such as servicing and
repairs may only be performed by the manufacturer
or by specialists expressly so authorized by the man-
ufacturer.
Maintain the following intervals:
Parts affected
– System components: e.g. carrying
systems, tube connections*
– Accessories
– Test bag
– Oxygen fittings
– Specified wear parts relevant to
safety
– Oxygen fittings
– Specified wear parts relevant to
safety
Steel and aluminum oxygen cylinders
*Disposable hose system WM 28110 does not
require any maintenance.
Person to carry
out
Manufacturer or
specialists expressly
so authorized by
the manufacturer