EN-16
Problem / Message
Not Executable
--------
Date/Time Not Set
----
Memory Error
---------
Memory full
---------
Load/Unload Weight
--
Unit change not
allowed!
Out of range
---------
Possible cause
Instrument cannot
------
execute printer
instruction.
Printer error
---
messages.
---------
Statistics or
----------
Totaling Memory is
full.
No sample or the
---
same sample is on
the weighing pan.
Changed unit
during Statistics or
Totaling
application.
Sample weight is
---------
out of range
SF40A Printer
Correction
Check the balance.
Check the instrument.
Reset the printer.
Set date/time
Back-up battery
depleted
Please contact OHAUS
service.
Start a new Statistics or
Totaling application.
Put a new sample or
remove it and put it
again on the pan.
Reset the balance.
Change the unit back to
the first one.
Sample weight must be
in the range of 70% -
130% of the current
average value.