Battery Replacement - Benning CM 3 Instrucciones De Servicio

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
Any work to be carried out on an opened BENNING CM 3 under voltage is
strictly reserved for qualified electrotechnical personnel who must take special
precautionary measures to avoid accidents.
This is how to ensure that the BENNING CM 3 is free from any voltage before
the instrument is opened:
-
first remove the BENNING CM 3 from measured object.
-
push the selector switch  to the "OFF" position.
9.1 Instrument safe­guarding
In certain circumstances, safety during the use of the BENNING CM 3 can no
longer be guaranteed; when for instance:
-
there is visible damage to the instrument
-
measurement errors occur
-
the instrument has been subjected to prolonged storage under unfavorable
conditions and
-
the instrument has been subjected to severe transport stresses.
In such cases the BENNING CM 3 must be immediately switched off, removed
from the measurement points and secured against any future unintentional
operation.
9.2 Cleaning
Wipe the case of the BENNING CM 3 with a clean dry cloth (exception:
special cleaning cloths). Do not use any solvents and/or abrasives to clean the
BENNING CM 3.

9.3 Battery replacement

Remove test leads and turn the power off before opening the
BeNNING CM 3! Dangerous voltage!
The BENNING CM 3 is powered by one 9-V block battery. Battery replacement
(see figure 4) becomes necessary when the low battery indicator appears in
the display .
This is how to change the batteries:
-
Remove the BENNING CM 3 from the measured circuit.
-
Place the BENNING CM 3 on its front side and remove the screw from the
battery box lid on the case back.
-
Slide the battery lid sideways from its position.
-
Remove the empty battery from the battery box and carefully disconnect the
battery leads from the battery.
-
Snap the battery connector leads to the terminals of a new battery in the
correct direction and place the leads so that they do not become pinched
between the case back and case front. Reinsert the battery into the battery
box.
-
Replace and fasten the battery box lid with the screw. Ensure that the
battery lead does not become pinched by the lid.
see Figure 4:
Please contribute to environmental protection! Batteries should
not be thrown into domestic refuse bins! They can be discarded
at collection points for old batteries or special refuse. Please
contact your municipality for more information.
9.4 Calibration
To maintain the specified precision of the measurement results, the instrument
must be recalibrated at regular intervals by our factory service. We recommend
a recalibration interval of one year. Send the appliance to the following
address:
Benning Elektrotechnik & Elektronik GmbH & CO. KG
Service Centre
Robert-Bosch-Str. 20
D - 46397 Bocholt
10. environmental notice
At the end of the product's useful life, please dispose of it at
appropriate collection points provided in your country.
06/ 2008
Battery replacement
BENNING CM 3
9