Guidance and manufacturer's declaration – electromagnetic immunity
The thermometer are intended for use in the electromagnetic environment specified
below. The customers or the users of these thermometer should assure that it is used
in such environment.
Immunity test
Conducted RF IEC
61000 -4-6
e
Radiated RF IEC
61000-4-3
IEC 60601
Compliance
Test level
level
Not applicable
3 V/m 80 MHz
3 V/m
to 2.5 GHz
Electromagnetic environment
– guidance
Portable and mobile RF
communications equipment
should be used no closer to
any part of the thermometer
including cables, than the
recommended separation
distance calculated from the
equation appropriate to the
frequency of the transmitter.
Recommend separation
distance
d = 1.2 √P
d = 1.2 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to he
transmitter manufacturer and d
is the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters as determined by
an electromagnetic site survey,
65