Light Guide; Care And Maintenance - Philips ZOOM! WhiteSpeed Manual Del Usuario

Ocultar thumbs Ver también para ZOOM! WhiteSpeed:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 12
1.4 Accelerator Assembly (refer to images in front)
1. Remove all items from shipping carton.
2. Snap the casters into the feet of the Accelerator Base
and lock the casters.
3. Place the Accelerator Base upright on the floor. Insert
the Accelerator Stand into the Accelerator Base by
aligning the pin in the Accelerator Stand with the slot
in the Accelerator Base and pushing the Stand securely
into place.
4. Assemble the lock washer with the big teeth facing each
other and place it on the bolt, followed by the Fender
(large) washer. Reaching under the Accelerator Base,
screw the bolt into the bottom of the Base, securing it to
the Accelerator Stand.
5. To attach the Accelerator Head, loosen the knobs by
the socket at the end of the Accelerator Boom by 4-5
turns. Gently, but firmly, push the Accelerator Head into
the socket. It will be held securely in place by the socket
springs.
6. Place the Accelerator Boom onto the Accelerator Stand
by carefully aligning the hole in the Boom with the hole
in the yoke. Insert the screw into the hole and tighten.
Keep tightening the screw until it is completely in the
pocket. The screw should be flush or inset into the
pocket. The screw has glue on the threads (yellow in
color) that releases as the screw is tightened. Therefore it
is important to finish tightening the screw upon insertion
or the glue may stop the screw from being fully inserted.
7. Connect the electrical cord from the Accelerator
Boom to the cords from the Accelerator Stand and the
Accelerator Head. The Accelerator head connector must
be turned until you feel it lock into place.
8. When you are ready to test the Accelerator, activate the
unit by turning on the main power switch on the right
side of the Power Pack.
1.5 Set Up
Plug the Accelerator power cord protruding from the
Accelerator Base directly into a properly rated wall outlet.
The Accelerator is now ready for use with a Philips
Zoom! Chairside Light-Activated Whitening Kit. Please
refer to the Kit's manual for further set up and get
application instructions.
5
*Based on the Vitapan Classical shade guide arranged in value order. Vita and Vitapan are trademarks of Vita Zahnfabrik, AG Germany.
2. Philips Zoom! Chairside Light-
Activated Whitening Treatment

2.1 Light Guide

The Light Guide is designed to ensure proper distance and
alignment to maintain correct light exposure on the smile
zone. Remove Light Guide from Philips Zoom! Chairside
Light-Activated Whitening Kit and place on end of Accelerator
Head. Turn on the Accelerator using power button located on
the side of the Power Pack. When the Light guide is properly
attached, the Power Pack display will read "Guide Status-Guide
Attached" followed by "X of X Sessions". Each new Light
Guide is loaded with four 15-minute activation sessions.
2.2 Align Accelerator Head and Light Guide
to Retractor
Once whitening gel application is complete, raise back of
chair so patient is in a slightly reclined position (45 degree
angle) and able to swallow comfortably while maintaining light
pressure on bite block. Slide Accelerator into place next to
patient. Accelerator Boom should be movable, but stable.
To align the Accelerator Head:
1. Grasp the Accelerator Head with one hand and bring it
close to the patient's mouth. Use your other hand to lift
or lower the Accelerator Boom as needed to make sure
Accelerator Head is aimed directly at the patient's teeth.
2. Use both hands to finalize positioning of Accelerator Head
so that Light Guide and IsoPrep Retractor slots are aligned.
Tighten Accelerator Head and Accelerator Boom knobs (4).
Lock casters in position. While engaged, the Light Guide
and IsoPrep Retractor will allow the Accelerator Head to
follow minor movements of the patient's head to ensure
proper alignment.
2.3 Running a Whitening Procedure Cycle
Each Philips Zoom! Chairside Light-Activated Whitening
Procedure consists of three 15-minute whitening sessions
with a pause between each session to allow the operator
to remove and replace the whitening gel. Each Light Guide
includes a fourth session for use on patients with shades A3
or darker.
1. Ensure the patient is properly prepared according to the
Kit manual with whitening gel and all protective barriers
in place.
2. Press the "Intensity" button to toggle through high, medium,
and low light intensity until the desired intensity is selected.
This setting can be changed at any time during the
whitening procedure.
3. Press the "Start" button to turn on light and activate
session timer.
4. Once light is on, a countdown will be displayed. The
Accelerator will beep once when 3 minutes are remaining
and beep three times in the final 3 seconds of the
15 minute session. When time reaches zero, light will turn
off and a long beep will sound.
5. Point out to the patient that progress of each session will
be tracked by 4 white lights on top of the Accelerator
Head, each indicating one quarter of a session.
6. If a session must be interrupted, press the "Pause" button.
The remaining session time will blink on the display. To
reactivate the session, press either the "Pause" or "Start"
button. DO NOT PRESS THE "STOP" BUTTON AT
ANY TIME DURING THIS PROCESS OR AN ENTIRE
15-MINUTE SESSION WILL BE LOST.
7. At the end of each 15-minute session, the display will
indicate the number of remaining sessions left on the Light
Guide. Press "Start" to active the light for the next session.
8. When all sessions on a Light Guide are used, the display
will read "Sessions Complete" for three seconds, followed
by "Guide Status – Replace Guide". Detach the Light
Guide from IsoPrep Retractor and pull Accelerator away
from patient.
Tip: Use High setting for maximum whitening, Low
setting for maximum heat and sensitivity
management. However, please note that lower light
intensity may reduce results.
Tip: After each session, make sure the patient's lips and
gums are consistently and thoroughly covered to
help prevent any soft tissue exposure to the LED
lights and any whitening gel.

3. Care and Maintenance

The Accelerator should be cleaned and disinfected after each
whitening session. Do not spray or pour liquids directly onto
the machine. Apply cleaners to a soft cloth or paper towel
before using. Do not use abrasive cleaners or solvents. The
following cleaners are recommended: Sanicloth, CaviWipes,
SciCan, Birex, BioSurf, Cidex and IPA. Follow your cleaning
products instructions for use to ensure proper cleaning of
the unit.
For optimum performance, ensure that the output window
on the Accelerator Head remains clean and free of particles
or streaks.
The Accelerator has no user-serviceable or user-replaceable
parts with the exception of two replaceable fuses that are
inside the Power Pack housing.
The Accelerator complies with relevant international and
national law and standards on EMC (electro-magnetic
compatibility) for this type of product when used as intended.
Such laws and standards define both the permissible
electromagnetic emission levels from product and its
required immunity to electromagnetic interference from
external sources.
Other electronic products exceeding the limits defined in such
EMC standards could, under unusual circumstances, affect the
operation of the product.
• Medical electrical products need special precautions
regarding EMC, and need to be installed and put into service
according to EMC information provided in this manual.
• The use of accessories and cables other than those
specified may result in increased emission or decreased
immunity levels.
• The product should not be used adjacent to other products
and that if adjacent use is necessary, it should be observed
to verify normal operation.
• To avoid the risk of electrical shock, do not remove any
System panels.
• Refer servicing to qualified personnel.
• Never operate the System with a damaged power cord.
Storage
The Accelerator Boom is designed to fold down to minimize
overall Accelerator footprint for convenient storage. To do so,
loosen Accelerator Boom knobs and fold Accelerator Boom
down and in towards the Accelerator Stand.
Disposal
Discus Dental is concerned about the preservation of the
environment. Please dispose of this product properly at the
end of its useful life according to local laws and regulations for
electronic equipment waste.