Guidance and declaration of manufacturer-electromagnetic emissions
The device is intended for use in the electromagnetic environment specified below. The customer or the user
of the device should assure that it is used in such an environment.
Emissions test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
N/A
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
N/A
IEC 61000-3-3
Electromagnetic environment – guidance
The device uses RF energy only for its internal
function. Therefore, its emissions are very low and
are not likely to cause any interference in
nearby electronic equipment.
The device is suitable for use in all establishments,
including domestic establishments and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Guidance and manufacturer's declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environment specified below. The customer or the user
of the device should assure that it is used in such an environment.
IEC 60601 test
Immunity test
Compliance level
level
Electrostatic
± 6 kV contact
± 6 kV contact
discharge (ESD) IEC
± 8 kV air
± 8 kV air
61000-4-2
Electrical fast
± 2 kV for
± 2 kV for power
transient/burst
power supply
supply lines
IEC 61000-4-4
lines
Surge IEC
± 1 kV line(s) to
± 1 kV line(s) to line(s)
61000-4-5
line(s)
<5% UT
(>95% dip
<5% UT
in UT)
(>95% dip in UT)
for 0.5 cycle
for 0.5 cycle
Voltage dips, short
40% UT
40% UT
interruptions and
(60% dip in UT)
(60% dip in UT)
voltage variations
for 5 cycles
for 5 cycles
on power supply
70% UT
70% UT
input lines
(30% dip in UT)
(30% dip in UT)
IEC 61000-4-11
for 25 cycles
for 25 cycles
<5% UT
<5% UT
(>95% dip
(>95% dip in UT)
in UT)
for 5 s
for 5 s
Power frequency
(50/60 Hz)
3 A/m
3 A/m
magnetic field
IEC 61000-4-8
Electromagnetic environment –
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the device requires
continued operation during
power mains interruptions, it is
recommended that the device
be powered from an uninterruptible
power supply or a battery.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.