Guidance And Manufacturer's Declaration-Electromagnetic Immunity
Immunity Test
IEC 60601 Test Level
Conduct RF
3 V 150 KHz to 80 MHz
IEC 61000-4-6
Radiated RF
3 V/m 80 MHz to 2,5 GHz
IEC 61000-4-3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fived transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
a
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fived RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the COMPAT
exceeds the applicable RF compliance level above, the COMPAT
performance is observed, additional measures may be necessary, such as reorienting or relocating the COMPAT
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
b
For Equipment and Systems That Are Not Life-Supporting
Compliance Level
Not Applicable
10 V/m
Standard Enteral Feeding Pump should be observed to verify normal operation. If abnormal
®
Electromagnetic Environment Guidance
Portable and mobile RF communications equipment should be used
no closer to any part of the COMPAT
Standard Enteral Feeding Pump,
®
including cables, than the recommended separation distance calculated
from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,17√P
d = 0,35√P 80 MHz to 800 MHz
d = 0,7√P 800 MHz to 2,5 GHz
where p is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in metres (m).
from fived RF transmitters, as determined by an electromagnetic site
survey,
should be less than the compliance level in each frequency
a
range.
Interference may occur in the vicinity of equipment marked
b
with the following symbol:
Standard Enteral Feeding Pump is used
®
Standard Enteral Feeding Pump.
®
EN
11