Audio Precision B Serie Instrucciones De Instalación página 53

Tabla de contenido

Publicidad

Characteristic
IMD Measurement
Test Signal Compatibility
SMPTE & MOD
DFD & CCIF
IMD Measured
SMPTE
MOD
DFD
CCIF
Measurement Range
Accuracy
1, 7
Residual IMD
SMPTE & MOD
DFD
Frequency Measurement
Range
Accuracy
Resolution
Phase Measurement
Ranges
1
Accuracy
Resolution
APx515 B Series Audio Analyzer: Specifications
Specifications
Any combination of 40 Hz–1 kHz (LF)
and 240 Hz–60 kHz (HF), mixed in
any ratio from 1:1 to 10:1 (LF:HF)
Any two-tone combination with mean
frequency of 250 kHz–60 kHz and a
difference frequency of 80 Hz–2.0 kHz
Amplitude modulation of HF tone.
d2, d3, d2+d3, or d2+d3+d4+d5
d2, d3, d2+d3, or d2+d3+d4+d5
d2 only
0 to 20%
±0.5 dB
<5 Hz to 0.499 • SR
±(0.0003% + 100 μHz)
6 digits
–90 to +270, ±180, or 0 to 360 deg
0.001 deg
Supplemental Information
LF tone must be 1/6 • HF tone.
F
= (F1 + F2)/2
mean
F
= |F2 - F1|
diff
must be  6 • F
F
mean
diff
Measurement BW is ≈40–750 Hz.
Use "d2+d3" for measurements per
IEC-60268.
Use "d2+d3" for measurements per
IEC-60268
"CCIF" is an archaic form of DFD that
measures only the d2 product. CCIF
uses a different 0 dB reference giving
readings 2x higher than DFD.
Typically <–140 dBFS
Typically <–150 dBFS
Typically 0.001 deg