36
Characteristic
GENERAL/ENVIRONMENTAL
Power Requirements
Temperature Range
Operating
Storage
Humidity
Max Operating Altitude
Stabilization Time
EMC
Safety
Dimensions (W x H x D)
Weight
Specifications
100, 120, 230 or 240 Vac, 50–60 Hz,
with safety ground via approved
power cord, 75 VA max.
0° C to +45° C
–40° C to +75° C
10 % to 80 %, non-condensing
3,000 m [9,840 feet]
20 minutes
Complies with Directive 2004/108/EC,
IEC 61326-1:2005, EN 61326-1:2006.
Radiated and conducted emissions
are within Class B limits of CISPR 11.
Complies with Directive 2006/95/EC,
IEC 61010-1:2001, EN 61010-1:2001,
CAN/CSA-C22.2 No. 61010-1-04, and
UL Std No. 61010-1 (2nd Edition).
370 x 86 x 343 mm
[14.55 x 3.40 x 13.53 in]
4.45 kg [9.8 lbs]
APx515 B Series Audio Analyzer: Specifications
Supplemental Information
Typical operating range is 90–110 Vac
(100V), 108–132 Vac (120V), 198–242
Vac (230V), or 216–264 Vac (240V).
Allow up to 1 hour per 10°C if unit has
been exposed to a significant change
in temperature. Allow 24–48 hours to
recover if condensation has occurred.
Emission and immunity levels are influ-
enced by the shielding performance of
interface and signal cables attached to
the instrument. EMC compliance was
demonstrated using Audio Precision
cables
Equipment Class I,
Installation Category II,
Pollution Degree 2,
Measurement Category I
2U rack mount tray available. D is
351 mm [13.56 in] if rear panel option
keys are installed

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Apx515

Tabla de contenido