APPENDIX B: Error Codes
Error Code
Description
Error 1
Pre-warm bypass flow error
Error 2
Pre-warm inlet pressure error
Error 3
Pre-warm nominal flow error
Error 4
Inlet pressure not stable
Error 5
Inlet pressure out of range
Error 11
Patient temperature 1 out of range at low test temperature (10 °C)
Error 12
Patient temperature 1 out of range at mid-low test temperature (25 °C)
Error 13
Patient temperature 1 out of range at mid-high test temperature (35 °C)
Error 14
Patient temperature 1 out of range at high test temperature (40 °C)
Error 15
Patient temperature 1 out of adjustment range limit
Error 16
Patient temperature 1 measurement unstable
Error 21
Patient temperature 2 out of range at low test temperature (10 °C)
Error 22
Patient temperature 2 out of range at mid-low test temperature (25 °C)
Error 23
Patient temperature 2 out of range at mid-high test temperature (35 °C)
Error 24
Patient temperature 2 out of range at high test temperature (40 °C)
Error 25
Patient temperature 2 out of adjustment range limit
Error 26
Patient temperature 2 measurement unstable
Error 30
Temperature out and patient temperature 2 difference out of range
Error 31
Temperature out measurement unstable
Error 32
Temperature out and simulator difference out of range
Error 40
Water temperature 1 is out of range at 6 °C
Error 41
Water temperature 1 is out of range at 28 °C
Error 42
Water temperature 1 is out of range at 40.5 °C
Error 43
Water temperature 1 off of conversion chart table at 1 °C
Error 44
Water temperature 1 off of conversion table at 47 °C
Error 50
Water temperature 2 is out of range at 6 °C
Error 51
Water temperature 2 is out of range at 28 °C
Error 52
Water temperature 2 is out of range at 40.5 °C
Error 53
Water temperature 2 off of conversion chart table at 1 °C
Error 54
Water temperature 2 off of conversion table at 47 °C
Error 60
Water temperature 3 is out of range at 6 °C
Error 61
Water temperature 3 is out of range at 28 °C
Error 62
Water temperature 3 is out of range at 40.5 °C
Error 63
Water temperature 3 off of conversion chart table at 1 °C
Error 64
Water temperature 3 off of conversion table at 47 °C
Error 70
Water temperature 4 is out of range at 6 °C
Error 71
Water temperature 4 is off of conversion table at 1 °C
Error 80
Water check time out - Unable to reach calibration temperature
Error 81
Water check time out - Difference between water temperature and reference temperature is greater than 2 °C
Error 82
Water check time out - other condition
Error 90
Heater test- flow is below 1 liter per minute
Error 91
Heater test- element 1 failure
Error 92
Heater test- element 2 failure
Error 93
Heater test- element 3 failure
Error 94
Heater test- element 4 failure
Error 95
Heater test- multiple elements fail
Error 99
Patient temperature out of calibration- no control
Error 100
Unable to initiate user diagnostic mode
Error 101
Unable to set calibration parameters
Error 102
Water level not full at start of calibration
Error 103
Switch settings incorrect
Error 104
Unknown error at startup
Error 105
Non recoverable navigation error
Error 106
Unknown error during pre-warm