Cushion Valve Adjustment
CLOCKWISE TO
TERMINAL IMPACT
User Instructions should be supplied to the wearer.
The user shall be aware that changes or modifications not expressly approved could void the operating
licences and exemptions.
This device complies with part 15 of the FCC Rules.
Operation is subject to the following two conditions:
2.
This device may not cause harmful interference.
3.
This device must accept any interference received, including interference that may cause undesired
operation.
Maintenance
In normal service an Endolite limb should be inspected annually to check for correct function, lubrication and security of
assembly.
Adjustment and maintenance must be carried out by competent, trained personnel.
If in doubt contact Blatchford Products Limited.
Should unit require return to Blatchford, please remove and dispose of the batteries.
(Disposal as per local waste handling regulations).
Cushion Valve
Adjust the Cushion Valve prior to commencement of programming.
ROTATE VALVE
For increased damping of terminal impact, rotate the valve
clockwise (2.5 A/F Wrench Key).
REDUCE
Adjust the cushion resistance whilst the amputee is walking slowly.
At this speed the damping level should be set to the highest
2.5
resistance that still permits the knee to extend fully and remain
extended during the entire stance phase. This should result in the
elimination of terminal impact at higher speeds.

Reset Switch

Initial Set Up, Calibration and Programming
1.
2.
Reset
3.
Switch
Location
Fine Tuning
1.
2.
Basic Reset
1.
Note: a basic reset should always be carried out after fitting
batteries.
When would you carry out any of the above resets?
selects factory default program settings
enters calibration mode followed by programming mode.
Press reset button, confirmed by 1 beep.
Within 2 seconds press reset button again, confirmed by 2
beeps.
Within 2 seconds press reset button again, confirmed by 3
beeps.
enters program mode starting with current program settings
Press reset button, confirmed by 1 beep.
Within 2 seconds press reset button again, confirmed by 2
beeps.
reverts to users programmed settings in normal operation
mode.
Press reset button, confirmed by 1 beep.
No response from the limb when trying any of the knee
activation sequences
Limb mode unknown
After changing or reconnecting batteries (Basic Reset)
3
938250/4-0719