Ambu Mark IV Resuscitator Manual Del Usuario página 12

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
CAUTION
Avoid using substances containing phenol to clean the product. Phenol will
cause premature wear and degradation of the materials or reduced product life span.
Promptly remove all residue of cleaning materials from the resuscitator. Residues
may cause premature wear or reduced product life span.
The following steps are generally recommended. Select proper methods for the
resuscitator parts in question according to the table above.
8.5 Disinfecting and/or sterilising
Select heated or chemical disinfection according to local standards for disinfection
and the table of applicable methods.
Washing machine - heat-disinfecting program:
Automatic washing machines designed for medical equipment will normally
include program cycles for heat disinfection.
Autoclaving: Use standard autoclaving equipment adjusted for 134 °C (273 °F) for
resuscitator parts according to the table above.
Sterilisation by gravity steam autoclave running a full cycle at 134 degrees Celsius for
10 minutes at a standard pressure and then dry 30 minutes.
For drying and cooling leave the parts to dry and/or cool completely before reas-
sembling the resuscitator.
Note: that it is the user's responsibility to qualify any deviations from the recom-
mended method of processing
12
8.6 Inspection of parts
After cleaning, disinfecting and/or sterilising carefully inspect all parts for damage
and residuals or excessive wear and replace if necessary. Some methods may cause
discoloration of rubber parts without impact on their lifetime. In case of material
deterioration e.g. cracking, the parts should be replaced.
In case of autoclaving, the oxygen reservoir bag can look slightly wrinkled. This has
no impact on its life or function.
When inserting the valve housing of the inlet valve: Make sure that the bag opening
seats smoothly against the flange.
When inserting the patient valve disc: Make sure that the butt of the stem is pushed
through the hole in the middle of the patient valve housing.
When mounting the outlet cap (option): Note that the slot of the outlet cap should
face the patient connector.
8.7 Reassembly
Reassemble the parts of the resuscitator as shown in the illustration. Mark IV (7.1),
Mark IV Baby (7.2)
9. Test of function 
Resuscitator
Close the pressure-limiting valve with the override cap (This only applies to the ver-
sion with pressure limiting valve) and close the patient connector with the thumb
(8.1). Briskly squeeze the bag. The resuscitator shall offer resistance to the squeeze.
Open the pressure-limiting valve by opening the override cap or by removing the
finger and repeating the procedure. The pressure limiting valve should now be acti-
vated and it should be possible to hear the expiratory flow from the valve.

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Mark iv baby resuscitator

Tabla de contenido