Guidance and manufacture's declaration – electromagnetic immunity
FO3001 is intended for use in the electromagnetic environment specified below. The user
should assure that it is used in such an environment.
Immunity test
IEC 60601 test
level
Electrostatic
±8 kV contact
discharge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
Surge IEC 61000-
± 1 kV line(s) to
4-5
line(s)
Voltage dips,
<5% UT
short
(>95% dip in UT)
interruptions and
for 0.5 cycle
voltage variations
on power supply
40% UT
input lines IEC
(60% dip in UT)
61000-4-11
for 5 cycles
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
Power frequency
10V/m
(50Hz/60Hz)
magnetic field IEC
61000-4-8
NOTE
UT is the a.c. mains voltage prior to application of the test level.
Compliance
Electromagnetic environment -
level
guidance
±8 kV contact
Floors should be wood, concrete or
±15 kV air
ceramic tile. If floor is covered with
synthetic material, the relative
humidity should be at least 30%.
±2kV for power
Mains power quality should be that of a
supply lines
typical commercial or hospital
environment.
±1kV differential
Mains power quality should be that of a
mode
typical commercial or hospital
environment.
<5% UT
Mains power quality should be that of a
(>95% dip in UT)
typical commercial or hospital
for 0.5 cycle
environment. If the user of the DEVICE
requires continued operation during
40% UT
power mains interruptions, it is
(60% dip in UT)
recommended that the DEVICE be
for 5 cycles
powered from an uninterruptible
power supply or a battery.
70% UT
(30% dip in UT)
for 25 cycles
<5% UT
(>95% dip in UT)
for 5 sec
10V/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
35
Ficheiro No.: FSHF-FO3001
Version: A/1