Guidance and manufacturer's declaration – electromagnetic emission
The MiniElite aerosol therapy equipment with Battery is intended for use in the electromagnetic
environment specified below. The customer or the user of the MiniElite aerosol therapy equipment with
battery should assure that is used in such an environment.
Emission Test
RF Emission CISPR 11
RF Emission CISPR 11
Harmonic Emission IEC 61000-3-2
Voltage fluctuation/flicker IEC 61000-3-3
Compliance
Electromagnetic
environment – guidance
The MiniElite aerosol therapy
Group 1
equipment with battery uses RF energy
only for its internal function. Therefore,
its RF emissions are very low and are
not likely to cause any interference in
nearby electronic equipment.
The MiniElite aerosol therapy
Class B
equipment with battery is suitable for
use in all establishments other than
Not applicable
domestic, and may be used in
domestic establishments and those
Not applicable
directly connected to the public low-
voltage power supply network that
supplies buildings used for domestic
purposes, provided the following
warning is heeded:
Warning: This equipment/system is
intended for use by healthcare
professionals only. This
equipment/system may cause radio
interference or may disrupt the
operation of nearby equipment. It
may be necessary to take mitigation
measures, such as re-orienting or
relocating the MiniElite aerosol
therapy equipment with battery or
shielding location.
MiniElite Compressor with Battery
Recommended separation distances between portable and mobile RF communications
equipment and the MiniElite aerosol therapy equipment with Battery
The MiniElite aerosol therapy equipment with battery is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or the user of MiniElite
aerosol therapy equipment with battery can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communication equipment (transmitters) and the
MiniElite aerosol therapy equipment as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum
output power of
transmitter (W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
NOTE 1: At 80MHz and 800MHz, the separation distance for the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Page 2 of 8
Separation distance according to frequency of transmitter (m)
150kHz to 80MHz
80MHz to 800MHz
Not applicable
4 P
– – –
0.40
– – –
1.26
– – –
4.00
– – –
12.65
– – –
40.00
800MHz to 2.5 GHz
2.33 P
0.23
0.74
2.33
7.37
23.3