EN-7
Grad [Grad] – Select the readability displayed. Reducing the readability may be needed for approval. (SET 1d,
SET [1]d, SET 10d)
Global Reset [Gbl rst] – Resets all settings to factory default values. (RESET)
.mMode.
4.4 Mode Menu
[
Count Mode [Count] - [SET ON or OFF]
Average Piece Weight (APW) Optimization [aPwW.OPT] - (SET ON, SET OFF)
Percent Mode [PerCent] - (SET ON, SET OFF)
Dynamic Mode [Dyn] – [SET ON, OFF] – [Auto] – [SET ON, OFF]
Density Mode [Density] - [SET ON, OFF] - [Solid, Liquid] – [Water, Other] – [T 20.0C] – [01.00000] – [Sinker] –
[010.000 ml]
When [T 20.0C] – [01.00000], [010.000] occur, use No to change the number and Yes to accept the change.
.Unit.
4.5 Unit Menu
[
The Unit menu is used to enable or disable a specific unit. (SET ON, SET OFF) The unit is indicated by a small
character next to Unit in the display (g = grams). The default setting is Grams SET ON and all other units SET
OFF.
T-Units – When unit [t] displays; press Yes to show the T-Unit settings; SET OFF, SET TT (Taiwan Tael), SET TH
(Hong Kong Tael), SET TS (Singapore Tael), SET TO (Tola) or SET TI (Tical).
M-Units – When unit [m] is displayed, press Yes to show the M-Unit settings; SET OFF, SET MO (Momme) or SET
ME (Mesghal).
Custom Unit – Custom Unit (C) is used to create a unit of measure not provided with the balance. The Custom Unit
is defined by a factor, a multiplier (E) and a least significant digit (LSD). The balance will use this to convert grams
to a custom unit of measure. (Example: 1 gram = 0.257206 Avoirdupois Dram, using a 4100g x 0.01g balance)
To create a custom unit, press Yes when unit [c] is displayed.
Factor – The Factor (F) is a value from 0.1000000 to 1.999999. When the Factor is displayed, the first digit is
flashing. Press Yes to accept its value and activate the next digit, or No to edit. When editing, press No until the
desired value appears, then press Yes to accept. Repeat until all digits have been accepted. When the Factor
flashes on the display, press Yes to accept or No to re-edit. (Example: F = 0.257206)
Multiplier (E) – The settings are, [E 0] (Fx1), [E 1] (Fx10), [E 2] (Fx100), [E 3] (Fx1000), [E-3] (F/1000),
[E-2] (F/100), and [E-1] (F/10). Press No to display the next setting, Yes to accept. (Example: E = 0).
Note: The multiplier selections are limited when the capacity in grams is exceeded.
LSD – The Least Significant Digit (LSD) is the number of displayed divisions (d) by which the weight is
incremented. The values are 1d, 2d, 5d, 10d, 100d or 0.5d. Press No to go to the next setting, press Yes to accept.
(Example LSD = 1d) Note: LSD options are limited if the readability in grams is exceeded.
The example custom unit will display [0.26 C] when 1g is placed on the pan.
4.6 Print Menu [.Print.]
Output Stable [Out.Stab] – Data will only be sent when the Stable indicator is on. This setting works with manual
button pressing or continuous and interval Auto Print. (SET ON, OFF)
Auto Print [A.Print] – Data will be continuously sent when [Continu] is set. Interval [Inter] sends data every 1
to 3600 seconds. When Stable [Stable] will send data when the balance detects a new stable reading. This can
be a weight value only [LOad] or it can also include a stable zero [L+zero]. [OFF] disables automatic printing.
Content [CONteNt] – The content in the data transmission can be modified. Each of the following settings can be
set on or off. Number Only [numMber] will only send the numeric result. Balance ID [Bal. ID] will add the Balance
serial number for traceability purposes. Reference [refer] will add reference information relevant to the current
mode. GLP [GLP] will send additional items to allow proper documentation of laboratory results.
]
]
Pioneer
™