Beper P303MED001 Manual Del Usuario página 12

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
EN
Cleaning and maintenance
The probe window must be kept clean and dry, to ensure accuracy of reading; to clean
this extremely delicate part, use a soft cloth slightly dampened with a 75% isopropyl
alcohol disinfectant solution. Do not use abrasive detergents. Do not use chemicals other
than isopropyl alcohol to clean the probe window.
After cleaning, wait at least 10 minutes to allow the thermometer to dry completely before
using it. Use a soft, dry cloth to clean outer body and display.
Warning: do not immerse thermometer in water.
Store the thermometer in a clean, dry environment, away from dust, dirt and other conta-
minants, and away from direct sunlight.
Strong electromagnetic fields may compromise the correct operation of the thermometer.
It's better to store the thermometer in its original box when not in use.
Specifications
Measuring range
Measuring site
Reference body site
Operation mode
Laboratory accuracy
Display resolution
Measure time
Operating temperature range
Transport and storage tempera-
ture range
Clinical accuracy
Shock
Forehead mode: 34.0°C – 43.0°C (93.2 °F –
109.4 °F)
Object mode: 0°C – 100°C (32°F – 212 °F)
Forehead (in forehead mode)
Oral (this thermometer converts the forehead
temperature to display its "oral equivalent")
Forehead mode (adjust mode)
Forehead mode:
ã 0.2°C (0.4°F) during 35.0°C ~ 42.0°C (95.9°F
~ 107.6 °F) at 15°C ~ 35°C (59.0°F ~ 95.0°F)
operating temperature range
ã 0.3°C (0.5°F) for other measuring and opera-
ting temperature range
Object mode:
ã 4% or ã 2°C (4°F) whichever is greater
0.1°C or 0.1°F
approx 1 second
10°C ~ 40°C (50°F ~ 104°F), 15% ~ 85% RH,
non-condensing
Atmospheric pressure: 700 hPa ~ 1060 hPa
-25°C ~ 55°C (-13°F ~ 131°F), 15% ~ 95% RH,
non-condensing
Atmospheric pressure: 700 hPa ~ 1060 hPa
Clinical bias: -0.09°C (-0.16°F)
Clinical repeatability: 0.13°C (0.23°F)
Limits of agreement: 0.87°C (1.57°F)
Withstands drops from 3 feet (0,90 m approx)
12

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

T238

Tabla de contenido