Maintenance, cleaning and sterilization - Handpiece
Warnings
Do not immerse the handpiece into disinfection solutions or into ultrasonic baths. These solutions may contain che-
micals that may impede or freeze the handpiece drive mechanism. The handpiece shall NOT be disassembled by the
user.
Limitations of reprocessing
The handpiece is tested to withstand 250 cycles of sterilization in steam autoclave.
Additional information
It is important that the handpiece is cleaned and properly lubricated between uses.
At point of use
1. Decontamination
Cleaning
Manual
Automated
Disinfection
Automated
Chemical
Drying
All manuals and user guides at all-guides.com
Shut off air and water spray function in motor (if available)
Check for proper function before use.
Clean off blood and debris by using a non-metal brush (to avoid scrat-
ching the instrument) under running cold water and dry the instrument
with a soft cloth.
Automatic reprocessing device for dental handpieces.
Not recommended. See warning above.
95 º C
Without or with mild detergents, low alkaline.
Automatic reprocessing device for dental handpieces.
Not recommended. See warning above.
Let air dry in upright position or wipe off with dry cloth.
8