Maintenance, cleaning and sterilization - LTA tips
Warnings- Use only with dedicated dental handpieces and hand instruments.
Limitations of reprocessing- See inspection and function.
Additional information- Used detergents: always follow the manufacturer's instruction.
At the point of use-
Check for function.
1. Decontamination
Cleaning
Disinfection
Drying
Inspection and function
2. Sterilization
(at the clinic validated
processes as per ISO
17665-1:2006)
Storage
All manuals and user guides at all-guides.com
Manual
Chemical
B-type steam
At least 3 minutes at 134 º C
Chemklave
Dry heat
60 º C max
Remove clogged debris by scrubbing the diamond
coated side against a cleaning block.
95 º C
Without or with mild detergents.
Not recommended.
Let air dry.
Visually inspection for appearance and function.
Eliminate worn tips.
Pack in sterilization bag of paper/clear film (accor-
ding to applicable standards). Check correct sealing
of the bag.
Do not use.
Do not use.
Protect from dust and dirt.