6. During use
Always ensure that the sensor is clean and undamaged. Switch on
the thermometer with the "ON/MEM" button.
6.1 Forehead measurement
After a short self-test and two short beeps, the thermometer is ready
for measuring the temperature at the forehead/temple. The infrared
thermometer is in 'Forehead thermometer' mode. This is indicated
by the symbol
Note that
– the thermometer must have been at least 30 minutes in the room
in which the measurement is made,
– the measuring result may be falsified by physical activity, funda-
mentally increased sweat formation on the forehead, consumption
of vasoconstrictive medications and skin irritations,
– the forehead or temple must be free of sweat and cosmetics.
Apply the thermometer to the left or right temple
and press the SCAN button. Move the thermometer
to and over the temple. During the measurement
you can hear short beeps signalling that the
thermometer has found a new maximum tempe-
rature. The end of the measuring time is signalled
with a long beep. Release the "SCAN" button. You
can now read off the temperature measured. The measurement
normally takes from 5 to 8 seconds, but can take up to 30 seconds.
The thermometer signals with two short beeps and a no longer
flashing forehead symbol that the thermometer is ready for a new
measurement. If you continue to hold down the "SCAN" button after
the measurement, you will enter a special measuring mode, in which
the thermometer will not make a correct measurement. In this event,
allow the thermometer to switch off automatically after about 1
minute and then switch it on again.