7. Trouble shooting
error
message
Problem
Autotest in progress, device
not yeat ready for.
Large fluctuation in ambient
temperature.
Ambient temperature belowr
10 °C or above 40 °C.
(<50 °F, >104 °F).
Device no longer functioning
correctly.
(1) Forehead thermometer
mode: the temperature mea-
sured is higher than 42.2 °C
(108 °F).
(2) SCAN mode: the temper-
ature measured is higher than
80 °C (176 °F)..
(1) Forehead thermometer
mode: the temperature
measured is lower than 34 °C
(93.2 °F).
(2) SCAN mode: the tempera-
ture measured is lower than
-22 °C (-7.6 °F).
Autotest not successful.

8. cleaning, storage, disposal

Clean the sensor tip after every use. Use a clean cloth or cotton bud
that can be moistened with disinfectant, alcohol or warm water.
To clean the entire device, use a soft cloth slightly moistened with a
mild soapy solution. Water must never get into the device. If water
All manuals and user guides at all-guides.com
17
Solution
Wait until the forehead symbol
no longer flashes.
Store the device for at least 30
minutes in the room where the
measurement will be taken.
The ambient temperature
must be between 10 °C and
40 °C
(50 °F, 104 °F).
Take out battery for approx.
1 minute and insert again. If
the error message is repeated,
consult your dealer or custo-
mer service.
Operate the thermometer
only between the specified
temperature ranges. If neces-
sary, clean the sensor tip. In
the event of a repeated error
message, contact your dealer
or customer service.
Insert new battery.