Important instruction: fixation of the inspiratory and expiratory tubes
Ensure that the weight of the tubes does not pull directly on the
patie t'sàheadào ào àtheài te face. Use cushioning to support
the tubes.
The inspiration and expiration tubing should slope away from
the patient and the Miniflow
does not come into contact with the baby.
(7)
Regularly check the CPAP parameters and the fixation.
Important instruction: regular check of the fixation:
Check the fit and fixation of Miniflow
Check the prongs and masks regularly for accumulation of secretions to avoid obstruction.
4.
Ventilators which can be connected to Miniflow
Miniflow
®
can be connected to all ventilators that meet the following requirements:
The ventilator has to offer a CPAP or non invasive mode which:
Controls the flow and pressure in such a manner that a constant CPAP pressure is maintained in the tubing
o
system.
Is able to work without an external flow sensor.
o
Measures the pressure inside the ventilator or at an external T-connector in the tubing system.
o
Tolerates small leakages in the tubing system.
o
If used in combination with Medin circuits, the ventilator has to have 22mm male or 15mm female connectors for the
inspiratory and expiratory tubes. If used with other circuits, the size of the connectors to Miniflow
female.
®
Miniflow
can among others be combined with:
Maquet SERVO Serie (i, u und n)
SLE 4000, 5000 und 6000
Hamilton T1, C1, MR1, C2, C3, C6, G5, S1
Newport Medical e360 Ventilator:
For Newport Medical e360, use only Newport approved breathing circuits when you use with the Miniflow
Miniflow
®
device from the circuit when e360 Circuit Check is performed.
4.1
Bonnets, masks, prongs
All bonnets, masks and prongs can only be combined with the Miniflow
5.
Accessories
Miniflow
®
20 pieces
Prongs – 10 pieces single use
X Small
Small
Medium
Medium wide
Masks – 10 pieces single use
X Small
Small
Masks – 5 pieces single use
X Large
Bonnets – 10 pieces single use
User Manual
All manuals and user guides at all-guides.com
®
so that the water condensation
®
and the accessories regularly to avoid pressure sores.
®
REF 4000
REF 1200-01
REF 1200-21
REF 1200-02
REF 1200-32
REF 1200-08
REF 1200-04
REF 1200-07
®
nCPAP generators.
Large
Large wide
X large
Medium
Large
EN
®
must be 10 mm
®
. Remove the
REF 1200-22
REF 1200-33
REF 1200-03
REF 1200-05
REF 1200-06
10