Maintenance
How to calibrate the i500
Turn on the i500 and connect the device to the software.
Run Calibration Wizard from Menu > Setting > Calibration.
Prepare the Calibration Tool and i500
Turn the dial of the calibration tool to the position.
Put the Handpiece into the calibration tool.
Click "Next" to start calibration process.
When the handpiece is mounted in the correct position , the system will
automatically acquire the data at the position
When data acquisition is completed at the position
next position.
Repeat the steps for positions
2
When data acquisition is completed at the
automatically calculate and show the calibration result.
4.2 Cleaning and sterilization procedure
4.2.1
Reusable tip
The i500 reusable tip is the part of the scanner that is inserted into the patient's
mouth during scanning. The tip can be reused for a limited number of times, but
it must be cleaned and sterilized between patients to avoid cross-contamination.
Clean the tip manually using soap water and a brush. Inspect the mirror of the tip
after cleaning. If the mirror appears stained, smudged or has a milky haze, repeat
the cleaning process using a soft dish brush and soap water. Dry the mirror carefully
with apaper towel.
Put the tip in a paper sterilization pouch.
1
.
1
, turn the dial to the
1
~
and
position.
8
LAST
position, the system will
LAST
The pouch should be sealed. Make sure it is airtight by using either a self-adhesive
pouch or a heat-sealed pouch.
Sterilize the wrapped tip in an autoclave with the following conditions:
At 121℃(249.8°F) for 30 minutes and 15 minutes dry
»
Use an autoclave program that dries the wrapped tip before opening the autoclave.
CAUTION
The mirror in the tip is a delicate optical component. It should be handled with care
as its clean and undamaged surface is critical for scan quality. Be careful not to
or smudge it before scanning a patient.
Never autoclave a tip that is not wrapped as this will leave stains on the mirror
which cannot be removed. See the autoclave manual for more information.
New tips must be cleaned and sterilized / autoclaved before using them for the first
time.
The scanner tip can be re-sterilized up to 20 times and must be disposed afterwards
as described in the next section on Disposal.
Medit will not responsible for any damage including distortion, blacken etc at other
condition.
Intraoral Scanner i500
13