intensity TWIN STIM III Manual De Instrucciones página 24

Ocultar thumbs Ver también para TWIN STIM III:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

All manuals and user guides at all-guides.com
2) Set Cycle Time (Optional)
Cycle time is adjustable from 5 – 30 seconds. Only modulation
has this parameter setting. Press [S] button cycle to enter this menu,
and then press the [▲] and [▼] button to adjust the setting.
3) Set Timer
Press [S] button cycle to enter this setting. The treatment time is
adjustable from 1 – 60 minutes or Continuous. Press [▲] or [▼]
button control to adjust setting. You can set the timer to
"Continuous" mode by pressing the [▲] control when it shows
60 minutes. The output will be shut off when time is up.
4) Set Pulse Width
Pulse Width is adjustable from 50µs – 300µs. Press [S] button to
enter this menu, then press [▲] or [▼] button to adjust the setting.
5) Set Pulse Rate
Pulse rate is adjustable from 1 Hz to 150 Hz (0.5 Hz to 5 Hz for
Burst). Press [S] button cycle to enter this menu, and then press
[▲] or [▼] button to adjust the setting.
4.7.2 EMS Setting
Press the [S] button cycle to enter the setting
state. The settings can be adjusted
according to the following steps:
1) Set the Therapeutic Program
There are 3 programs in EMS therapeutic
mode available — Synchronous, Alternate and Delay,. The therapeutic
program can be selected by pressing the [▲] and [▼] button. When you
choose to [S] program, program [S] outside of the box will be flashing.
24

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido