5.5 Menu overview
The menu structure is based on the VDMA standard sheet
24574-1. The menu structure is hierarchic. The top menu level
contains the main menu items, e.g. Pres, b.EF, diA, E. Each
main menu has additional submenu items.
The menu items may vary depending on the device configura-
tion. Not all menu items described below will necessarily ap-
ply to your device. Press the
the configuration. A 4-digit code will appear, e.g.
With the 4 digits tsav meaning:
t: Model
s: Number of switch-
ing outputs
a: Number of ana-
logue outputs
v: Device installation
type
P__i
Configuration
Equipment configuration
____
Display
____
Measurement display
PrES
Pressotronik
Pressure main menu
b. E F
Basic EF
Basic Adv. Functions
Diagnostic
DiA
Diagnostics menu
Exit
E
Exit menu level
The individual menu items will not be shown if the option
does not apply. Example: With a=0 the menu items for config-
uring the analogue output does not apply. You can then skip
the description for this item.
The Pressotronik (Pres) main menu is used to configure the
settings for the switching outputs or the Analogue outputs (if
applicable).
The basic device settings can be changed. General settings can
be configured under Basic Settings Advanced Functions (b.EF).
These settings should be configured first, as they affect the
displays and settings for the individual menus. These settings
are e.g. the units used.
The Diagnostic (dia) menu further contains diagnostics op-
tions.
BX130003 ◦ 02/2021
key in display mode to open
P = pressure measure-
ment
2, 4 or 6
0 for 1
i = no differentiation
Toggle
____
Toggle display
Out 1
out1
Settings menu OUT1
out_
Out x
If applicable
Settings menu OUTx
An 1
Analog Out 1
If applicable
Settings menu Analog 1
E
Exit
Exit menu level
You will find the detailed illustration of the entire menu
structure at the end of these instructions.
6 Maintenance
Always observe the applicable safety- and operating regula-
tions when performing any type of maintenance. Please refer
to the original operator's manual on the included CD or online
at www.buehler-technologies.com for maintenance informa-
tion.
7 Service and Repair
Please refer to the original operator's manual on the included
CD or online at www.buehler-technologies.com for a detailed
description of the unit including information on
troubleshooting and repair.
7.1 Troubleshooting
Problem /
Possible cause
Malfunction
No display
• No supply
voltage
Error messages on the display:
Alternating between Err and Exxx: e.g.
Error
• Ambient temper-
001
ature too low
Error
• Ambient temper-
002
ature too high
Error
• Cable brake or
064
defective trans-
mitter
• Transmitter de-
fective
Error
• Measuring range
128
exceeded
• Second plug has
no supply
Error
voltage
1024
• Internal error
Action
• Check cable and
replace, if neces-
sary
⇔
• Maintain limits
• Maintain limits
• Replace transmit-
ter and/or cable
• Replace transmit-
ter
• Check the system
and transmitter,
maintain the max.
pressure
• Check the supply
voltage for the
second plug.
• Be sure to switch
on or plug in the
power supply for
the second plug
before or shortly
after the first.
• Please contact
customer service
Bühler Technologies GmbH
11