Electromagnetic Compatibility

Like other electrical medical equipment, the VisionPro 26" LED Display, and the VisionPro SYNK
26" Wireless LED Display require special precautions to ensure electromagnetic compatibility with
other electrical medical devices� To ensure electromagnetic compatibility (EMC), the display must
be installed and operated according to the EMC information provided in this manual� The display
has been designed and tested to comply with IEC 60601-1-2 requirements for EMC with other
devices�
Note: This
equipment is for use in a professional healthcare environment� It is not for use in the RF
shielded room of a medical electrical system for magnetic resonance imaging, where the intensity
of EM disturbances is high�
Note: This
equipment is not likely susceptible to interference from HF surgical instruments in the
Special Environment of being in close proximity to an active HF surgical instrument� In the case
that HF surgical interference is observed, adjust the separation distance of the equipment�
Warning
Caution
Guidance and Manufacturer's Declaration: Electromagnetic Emissions
The VisionPro 26" LED Display, and the VisionPro SYNK 26" Wireless LED Display are intended for use in the electromagnetic
environment specified below� The customer or the user of the display should ensure it is used in such an environment�
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC61000-3-2
Voltage Fluctuations/ flicker
emissions IEC61000-3-3
Guidance and Manufacturer's Declaration — Electromagnetic Immunity
The VisionPro 26" LED Display, and the VisionPro SYNK 26" Wireless LED Display is intended for use in the electromagnetic
environment specified below� The customer or the user of the display should ensure that it is used in such an environment�
When this device is connected with other electrical equipment, leakage
currents may be additive� To minimize total leakage current per patient,
ensure that all systems are installed according to the requirements of
IEC 60601-1-1�
Portable and mobile RF communications equipment may affect the
normal function of the display�
Do not use cables or accessories other than those provided with the
display, as this may result in increased electromagnetic emissions or
decreased immunity to such emissions�
If the display is used adjacent to or stacked with other equipment,
observe and verify normal operation of the display in the configuration
in which it will be used prior to using it in a surgical procedure� Consult
the tables below for guidance in placing the display�
Compliance
Electromagnetic Environment - guidance
Group 1
The displays use RF energy only for their internal function; therefore,
their RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment�
Class B
The displays are suitable for use in all establishments other than
domestic establishments and those directly connected to the public
Class D
low-voltage power supply network that supplies buildings used for
domestic purposes, provided the following warning is heeded:
Complies
Warning: This system is intended for use by health care professionals
only� This system may cause radio interference or may disrupt the
operation of nearby equipment� It may be necessary to take mitigation
measures, such as reorienting or relocating the system or shielding the
location�