Cleaning: it cannot be cleaned under running water
Use a soft, damp cloth; a neutral, non-abrasive detergent can be used.
Disinfectinon: Use a sponge or a soft cloth moistened with a cold disinfectant
solution (such as, for example, a solution with a sodium hypochlorite concentration not
exceeding 2%), following the manufacturer's instructions. Ensure that the solution is
compatible with the material of which the base unit is made.
Sterilisation: the sterilisation cycle is not provided.
INTERNAL CHAMBER, ATOMISER AND EXTERNAL CHAMBER
Cleaning: they can be washed with warm water (approx. 40°C) using detergent
(dosed according to the detergent manufacturer's indications) and rinsed under
running water. When all parts have been cleaned, they can be dried with a soft cloth.
Disinfection: after being washed, they can be immersed into a cold disinfectant
solution (such as for example a solution with a sodium hypochlorite concentration not
exceeding 2%), following the manufacturer's instructions. It is possible to disinfect the
accessories by boiling them in water for up to 10 minutes: the parts must not be in
direct contact with the bottom of the pot.
Sterilisation: they are compatible with the sterilisation cycle in an autoclave at 121°C
for at least 15 min. (max. 30 min.) for a maximum of 20 cycles.
MAINTENANCE
The cleaning, disinfection and sterilisation operations of the nasal shower components
allow the continuous control of their appearance and can therefore highlight the need
for their replacement.
Repeated cleaning, disinfection and sterilisation cycles may alter the colour or
aesthetics of the components; this does not affect the operation of the device.
Sterilisation cycles can compromise the characteristics of the nasal shower material; it
is therefore advisable to check the appearance and integrity of the components after
each sterilisation.
The device must be replaced if damaged.
In case of prolonged non-use, it is recommended to recharge the battery every 6
months to ensure maximum battery efficiency.