Bair Hugger Model 750 Temperature Management Unit Operator's Manual

Viewing the Temperature Mode Timer

The temperature mode timer records the amount of time the temperature management
unit has been operating in the selected temperature mode� The timer resets itself every time
you select a different temperature mode�
To view the temperature mode timer, press the button for the current temperature mode,
and hold it for 2 seconds� The alphanumeric display will show the temperature mode time,
then go back to showing the temperature mode setting�

What to Do in Case of an Over-Temperature Condition

If an over-temperature condition occurs, the red Over-temp indicator light flashes and
an alarm sounds� The unit automatically turns the heater, blower, and operating mode
indicator lights OFF� The control panel will not respond to commands, but pressing any
button will silence the alarm�
-
if an over
temperature condition occurs
1�
Discontinue all temperature management therapy� If you are using the 241 blood/fluid
warming system, immediately stop fluid flow and discard the blood/fluid warming set�
2�
Unplug the temperature management unit�
3�
Contact a qualified service technician�

What to Do in Case of a Fault Condition

The Model 750 temperature management unit's software recognizes several non-hazardous
conditions and reports those conditions as faults� When a system fault occurs, the software
stores the fault code in memory, the amber Fault indicator light flashes, and an alarm
sounds� The control panel will not respond to commands, but pressing any button will
silence the alarm�
if a fault condition occurs
1�
Unplug the temperature management unit and wait for 5 minutes�
2�
Reconnect the temperature management unit to a grounded power source� The unit
will perform the normal power-on-reset sequence and then enter the Standby mode�
3�
Reselect the temperature setting�
4�
If the unit does not return to normal operation, contact a qualified service technician�
:
:
15