Important Information about the
Bair Paws Model 875 Warming Unit
Indications/Intended Use
The Model 875 warming unit is indicated for patient warming� This warming unit has been
designed for use with the Bair Paws warming gowns in all preoperative and postoperative
settings�
Contraindication
Do not apply heat to lower extremities during aortic cross-clamping� Thermal injury may
occur if heat is applied to ischemic limbs�
Warnings
Use only Bair Paws warming gowns with this warming unit� This warming unit has
•
been designed to operate safely with Bair Paws warming gowns; use with other products
may cause thermal injury� To the full extent permitted by law, the manufacturer and/or
importer declines all responsibility for thermal injury resulting from the unit being used
in conjunction with products other than Bair Paws warming gowns�
•
Do not continue warming if the red Over-temp indicator light illuminates or an audible
•
alarm sounds; thermal injury may result� Turn the warming unit off�
Do not use a forced-air warming device over transdermal medications; increased drug
•
delivery, patient death, or injury may occur�
Do not allow the warming unit's hose to contact the patient's skin during patient
•
warming� Thermal injury may result�
Equipment not suitable for use in the presence of a flammable anaesthetic mixture with
•
air or with oxygen or nitrous oxide�
This warming unit is not intended for use in the operating room�
•
Position the temperature controller cord and the hose away from the patient's neck or
•
shoulders to avoid entanglement and/or injury�
2
Do not warm patients with the warming unit's hose alone� Thermal injury
may result� Always connect the hose to a Bair Paws warming gown before
providing patient warming�
Bair Paws® Model 875 Warming Unit Operator's Manual