Technical Specifications - HoMedics HTD8216C Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 89

TECHNICAL SPECIFICATIONS

Measurement unit
Operating mode
Measuring site
Reference Body Site
Rated output range
Output range
Laboratory Accuracy
Display Resolution
Three-colour Backlight
(Colour Alarm)
Auto Power Off Time
Measuring Time
Memories
°C/°F
Adjusted mode (Body mode)
Direct mode (surface mode)
Forehead / Ear
Axillary
35.0-42.0°C :±0.2°C / 95.0-107.6°F:±0.4°F
Body mode: 34.0-42.9°C / 93.2-109.2°F
Surface mode: 0°C - 100.0°C / 32°F -212°F
Body mode:
34.0 - 34.9°C: ± 0.3°C / 93.2 - 94.8°F: ± 0.5°F
35.0 - 42.0°C: ± 0.2°C / 95.0 - 107.6°F: ± 0.4°F
42.1 - 42.9°C: ± 0.3°C / 107.8 - 109.2°F: ± 0.5°F
Surface mode:
± 2°C / ± 3.6°F
0.1°C/0.1°F
34 - 37.3°C / 93.2 - 99.1°F: Green (Normal Temperature)
37.4 - 38.0°C (Alarm point) / 99.3 - 100.4°F: Yellow (Slight Fever)
38.1 - 42.9°C / 100.6 - 109.2°F: Red (High Fever)
Note: Colour Alarm is only for Body mode, Surface mode is always with
Green backlight.
≤35s
≤ 5s (under ear mode)
≤2S(under Forehead mode)
3-10s (under Forehead scan mode)
1
GB | 7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Te-300-eu 1

Tabla de contenido