• There are no electrical connections from
the tonometer to the patient.
• The device has B-type electric shock
protection.
• Operation environment:
• Temperature: +10 °C to +35 °C
• Relative humidity: 30 % to 90 %
• Atmospheric pressure: 800 hPa-1,060 hPa
• Storage environment:
• Temperature: -10 °C to +55 °C
• Relative humidity: 10 % to 95 %
• Atmospheric pressure: 700 hPa-1,060 hPa
• Transport environment:
• Temperature: -40 °C to +70 °C
• Relative humidity: 10 % to 95 %
• Atmospheric pressure: 500 hPa-1,060 hPa
• Mode of operation: continuous.

ACCESSORIES

SKU
PRODUCT DESCRIPTION
104
BOX OF 100 PROBES
102
BOX OF 100 PROBES (ONLY USA)
7210
PROBE BASE COLLAR
540
PROBE BASE
550
TABLE STAND
560
WRIST STRAP
500
ALUMINIUM CASE
7171
BATTERY COVER & SCREW
7000
THREAD PLATE
543
PROBE BASE CLEANING CONTAINER
561
SILICON GRIP - WHITE
562
SILICON GRIP - PINK
563
SILICON GRIP - GREEN
564
SILICON GRIP - BLUE

MAINTENANCE

Follow local regulations and recycling
instructions regarding the disposal or recycling
of the Icare tonometerand accessories.
WARNING
The tonometer should only be opened by
qualified service personnel. It contains no
userserviceable parts, apart from the batteries
and a probe base. The Icare tonometerrequires
no routine servicing or calibration other than
changing the batteries at least every 12 months
or changing or cleaning the probe base. If
servicing is necessary, contact qualified service
personnel or your local Icare representative.
All manuals and user guides at all-guides.com

PERFORMANCE DATA

The performance data is obtained from
a clinical study, performed according to
American National Standard ANSI Z80.10-
2003 and International Standard ISO 8612.2
for tonometers. The study was performed ain
the Department of Ophthalmology, Helsinki
University Central Hospital. In the study,
158 patients were measured. The mean paired
difference and standard deviation (Goldmann-
Icare) were -0.4 mmHg and 3.4 mmHg.
A scattergram and Bland-Altman plot of the
results is shown in figure 14.

REPLACING THE PROBE BASE

Replace the probe base every twelve months.
Clean or replace the probe base if the error
messages E01 or E03 are displayed.
Instructions for replacing the probe base:
• Turn off the tonometer.
• Unscrew the probe base collar and put it in a
• Remove the probe base by tilting the
• Insert a new probe base into the tonometer.
• Screw the collar in, to lock the probe base.
WEIGHT DIMENSIONS
89,10 g
5,3 x 10,3 x 3,6 cm
89,10 g
5,3 x 10,3 x 3,6 cm
1g
11 mm x 10 mm
4g
7 mm x 38 mm
52g
73 mm x 52 mm x 38 mm
3g
270 mm x 10 mm x 10 mm
700g
240 mm x 280 mm x 72 mm
4g
42 mm x 17 mm x 13 mm
1g
4 mm x 10 mm x 3 mm
3g
5,6 cm x 2 cm
26g
135 mm x 46 mm x 23 mm
safe place.
tonometer downwards and use your fingers
to pull the probe base out of the tonometer.