Switching Off The Unit - Medisana 51085 Instrucciones De Manejo

Esfigmomanómetro de brazo mtx connect
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 92
GB
4 Operation
4.5
Troubleshooting
4.6
Switching off
the unit
30
3. First hold down the START button
together for more than 5 seconds, the MTX Connect switches to transfer
mode. The display shows the selected user memory, the time and the
Bluetooth symbol.
4. In the VitaDock App press on MTX BT - screen C to connect the two devices
and transfer the stored measurement values. The MTX Connect connects
with your smartphone and starts to transfer the stored measurement values.
The data transfer symbol
5. Once the measurement values have been transferred to the VitaDock App,
press the START button
not press any buttons for approx. 1 minute, the device automatically switches
to standby mode.
Error messages
Possible causes and solutions
Err 1
Measurement of the systolic blood pressure was completed.
Either the connection with the hose has been interrupted or no pulse
could be determined.
Check the connections between the cuff and the main unit.
Err 2
Non-physiological pressure has been detected.
Cause: the arm was moved during the measurement.
Repeat the measurement, keeping your arm still.
Err 3
If the cuff takes too long to inflate, it is possible that the cuff is
incorrectly positioned or that the connection with the hose is not airtight.
Check the connections and repeat the measurement.
Err 5
An average value could not be calculated due to unstable conditions
while the readings were being taken.
Repeat the measuring process, keeping your arm still and not speaking.
Err 6
There were too many errors during the measurement in MAM mode,
making it impossible to obtain a final result.
Read through the checklist for performing reliable measurements and
then repeat the measurement.
HI
The pressure in the cuff is too high (over 300 mmHg) or the pulse is too
high (over 200 beats per minute).
Relax for 5 minutes and repeat the measurement.
LO
The pulse is too low (less than 40 beats per minute).
Repeat the measurement.
If no buttons are pressed for approx. 1 minute, the device automatically switch-
es to standby mode. You can also switch to standby mode by pressing the
START button
. The blood pressure meter functions are now switched off.
The time and the selected user memory are continuously shown on the display.
The alarm function remains active.
If you do not want to use the time display and the alarm function, you can also
completely switch off the device by sliding the ON/OFF switch
of the device into the " OFF " position.
and then the MEMO button
flashes.
to switch off the MTX Connect again. If you do
on the back

Publicidad

Tabla de contenido
loading

Tabla de contenido