6.2 Applicator
A trigger button is located on the applicator handpiece. The flash lamp is activated by the trigger but-
ton.
The cartridge is located in the interior housing of the applicator.
The indicator lamp is located at the top centre of the applicator head and illuminates as follows:
Indicator lamp not illuminated
Indicator lamp illuminated green
Indicator lamp illuminated red
Indicator lamp illuminated orange
Indicator lamp flashes red/energycontrol lamp
flashes green
6.3 Cartridge
The beurer SalonPro System cartridge features a built-in skin colour sensor.
Using light-based hair removal on darker skin tones can have adverse consequences, such as burns,
blisters or discolouration of the skin (hyper- or hypopigmentation). To prevent such application errors,
the skin colour sensor in the beurer SalonPro System measures the colour of the treated skin before
each light pulse. If the skin colour sensor detects that the skin is too dark for the application of the beu-
rer SalonPro System, the device automatically ceases to emit pulses.
If no light pulse can be seen and the indicator lamp is illuminated red, this indicates that the colour of your
skin measured by the skin colour sensor is too dark to perform the treatment safely.
The beurer SalonPro System cartridge also features a skin contact sensor to protect the eyes. The sen-
sor was developed so that no light pulses can be emitted when the applicator is held in the air. If no
light pulse can be seen and the indicator lamp is illuminated orange, this indicates that the light area
on the cartridge does not have sufficient contact with the skin. Test the device on another part of your
body or contact beurer SalonPro System Customer Services.
The beurer SalonPro System cartridge features built-in UV protection, blocking dangerous UV radiation.
Skin colour sensor/skin contact sensor
Light area
Pulse is not ready.
Pulse is ready.
The pulse is blocked because the skin colour is
unsuitable (see "Hair and skin colour chart" on
the back page of these instructions for use).
The pulse is blocked because the device is not in
full contact with the skin.
Cartridge is not inserted properly.
30