Topcom TH-4654 Manual De Usuario página 13

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 50
The measurement
is not accurate or if
there is any doubt of
the measure result
The measurement
is not accurate or if
there is any doubt of
the measure result.
EAR & FOREHEAD THERMOMETER TH-4645/55
Please check if the
infrared sensor is
clean or not.
Please check the
measuring way is
correct or not.
Please check if the
thermometer and
patient gets
stabilized in the
room for 30 minutes
at least.
Are you using the
thermometer
indoor.
Please check if you
hold the
thermometer in
your hand for too
long this could
affect the accuracy.
Please check if the
low battery icon is
on the display
Clean the infrared sensor with
cotton swab according to user
the user manual.
Ensure you have read the
manual and know how to use
the thermometer properly.
Please keep the thermometer
and patient in the measuring
room at least 30 minutes
before using.
Please take the measurement
indoor.
Put the thermometer on the
table in the room where the
measurement is taking place
and let it cool down first.
Replace the battery.
13

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Th-4655

Tabla de contenido