R&S
Scope Rider RTH
Shock hazard caused by high voltages
To avoid electrical shock and personal injury, and to prevent damage to the
instrument or any other products connected to it, observe the following:
● Do not apply input voltages above the rating of the instrument and the
accessories.
● Use only probes, test leads, and adapters that comply with the mea-
surement category (CAT) of your measurement task.
● Test leads and measurement accessories used for multimeter measure-
ments on a live mains circuit must be rated for CAT III or CAT IV
according to IEC 61010-031. The voltage of the measured circuit must
not exceed the rated voltage value.
Maximum input voltage:
● At BNC inputs: CAT IV 300 V
● With probe R&S RT-ZI10 or R&S RT-ZI11: CAT IV 600 V,
CAT III 1000 V
● Meter input: CAT IV 600 V; CAT III 1000 V
Voltage ratings: V RMS (50 to 60 Hz) for AC sine wave and V DC for DC
applications.
Risk of electrical shock or fire
Voltages higher than 30 V RMS or 42 V peak or 60 V DC are regarded as
hazardous contact voltages. When working with hazardous contact vol-
tages, use appropriate protective measures to preclude direct contact with
the measurement setup:
● Use only insulated voltage probes, test leads and adapters.
● Do not touch voltages higher than 30 V RMS or 42 V peak or 60 V DC.
Getting Started 1326.1561.02 ─ 05
Instrument Tour
Top View
24