9. Maintenance, accessories / 10. Trouble shooting
Accessories
For details about the accessories (e. g. connectors), please refer to WIKA's price list, WIKA's product catalog on CD or or
contact our sales department.
10. Trouble shooting
Warning!
Open pressure connections only after the system is without pressure!
Take precautions with regard to remaining media in removed transducers. Remaining media in the
pressure port may be hazardous or toxic!
Remove the transducer from service and mark it to prevent it from being used again accidentally, if it
becomes damaged or unsafe for operation.
Have repairs performed by the manufacturer only.
Do not insert any pointed or hard objects into the pressure port for cleaning to prevent damage to the
diaphragm of the pressure connection.
Please verify in advance if pressure is being applied (valves/ ball valve etc. open) and if the right voltage supply and the right
type of wiring (2-wire/3-wire) has been chosen?
Failure
Output signal unchanged after change in
pressure
No output signal
No/False output signal
Abnormal output signal or
Abnormal zero point signal
WIKA Operating instructions/Betriebsanleitung/Mode d'emploi/Instrucciones de servicio WU-1X
Possible cause
Mechanical overload through overpressure
Wrong supply voltage or current spike
No/incorrect voltage supply or current spike
Cable break
Incorrectly wired (e.g. Connected as 2-wire
instead of 3-wire system)
Zero point set wrongly
Procedure
Replace instrument; if failure reoccurs, consult
the manufacturer
*)
Replace instrument
Adjust the voltage supply
Check connections and cable
Follow pin assignment
Adjust zero point correctly (see chapter 8); a
sufficiently accurate current/volt meter should
be used