Supplemental Oxygen - DeVilbiss Healthcare DV53 SLEEPCUBE STANDARD PLUS CPAP Serie Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Approximate Battery Run Time (with fully charged battery) @ 12 cm H
DC Power
Source/Battery
Battery Specs
Additional
Equipment Needed
DeVilbiss Cables
Required (refer to
DV51D-619
Accessories)
Heater Humidifier
Powered?
Approximate Run
Time
NOTE–Run times are at a breath rate of 20 breaths per minute, standard mask leak, and with heater setting of 8 where
applicable.
NOTE–The heater for the SleepCube's humidification system will not operate directly from a 12V DC power source. If you
need humidification, non-heated humidification is still available. Alternately, instead of connecting12V DC power directly to
the SleepCube, you can pass the DC power through an inverter that will deliver AC power to the system. This will allow the
heater to function. The power rating of the inverter must be at least 200 watts @ 110V~ / 400 watts @ 220V~.
NOTE–The SleepCube will automatically switch power sources between AC and 12V DC if both sources are available. If
your unit is connected to both AC and 12V DC power sources, it will operate on AC power as long as AC power is present
and automatically switch to 12V DC power if AC power is lost. When AC power is restored, it will return to the AC source.
You do not need to reset or adjust any settings when the power source switches, but remember that your heater will not
operate on 12V DC.

SUPPLEMENTAL OXYGEN

WARNING-OXYGEN USE
Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with
supplemental oxygen. Do not use this device near hot objects, volatile substances or sources of open flames.
Always turn on the SleepCube before turning on the oxygen source. Turn off the oxygen source before turning off the
SleepCube.
This device is not suitable for use in the presence of a flammable anesthetic mixture.
At a fixed flow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure
setting, patient breathing pattern, mask selection and leak rate. This warning applies to most types of CPAP devices.
Never allow the oxygen source to run continuously while connected to the SleepCube if the device is not in use. If
the SleepCube is not being used, turn off the oxygen flow.
CAUTION
Oxygen is prescribed gas and should only be administered under the supervision of a physician.
The flow setting of the oxygen source must be specified by a physician.
Supplemental oxygen is sometimes prescribed in addition to a CPAP for treatment of Obstructive Sleep Apnea. If your
physician prescribed oxygen for your sleep therapy, you can add the oxygen in one of two ways:
1.
Obtain the optional oxygen adapter (part #7353D-601) from your homecare provider and connect it to the air supply
port on the SleepCube. Connect one end of the air supply tubing to the open end of the adapter and the other end of
the air supply tubing to the mask.
2.
If your mask is equipped with an oxygen port, your physician may request that you apply oxygen directly to your mask.
EN - 10
SleepCube Connected To Vehicle
DC Car Power Source (Cigarette Lighter)
12 V
12V
120V~
Modified Sine
Wave Inverter
with 200 Watt
capacity
DV51D-696
NO
YES
5 nights
1 night
SleepCube Connected To Battery
12V
80 Amp-Hours
230V~ Modified
Sine Wave
Inverter with
400 Watt
Capacity
DV51D-619
DV15D-696
DV51D-696
YES
NO
1 night
7 nights
O setting
2
Deep Cycle Marine Battery
120 Amp Hour
120 VAC
Modified Sine
Wave Inverter
with 200 Watt
capacity
DV51D-696
YES
1 night
240 Amp Hour
230 VAC
Modified Sine
Wave Inverter
with 400 Watt
Capacity
DV51D-696
YES
1 night
SE-DV53-1

Publicidad

Tabla de contenido
loading

Tabla de contenido