DeVilbiss Healthcare SleepCube DV55 Serie Manual De Instrucciones página 17

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Emissions Test
Compliance
RF Emissions
Class B
CISPR 11
Harmonics
Class A
IEC 61000-3-2
Flicker
Complies
IEC 61000-3-3
Immunity Test
IEC 60601 Test Level
Electrostatic
±6kV contact
Discharge (ESD)
±8kV air
IEC 61000-4-2
Electrical Fast
±2kV on AC Mains
Transient/burst
±1kV for I/O Lines
IEC 61000-4-4
Surge
±1kV Differential
IEC 61000-4-5
±2kV Common
Voltage dips, short
>95% Dip for 0.5 Cycle
interruptions and
60% Dip for 5 Cycles
voltage variations
on power supply
30% Dip for 25 Cycles
input lines
>95% Dip fo 5 Seconds
IEC 61000-4-11
Power Frequency
50/60Hz Magnetic
3A/m
Field IEC 61000-4-8
Conducted RF
3 Vrms from
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
For transmitters rated at a maximum output power not listed above, the recommended separation distance D in meters
(m) can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
SE-DV56-1
Electromagnetic Enforcement – Guidance
This device is suitable for use in all establishments including domestic,
and those directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Compliance Level
±6kV contact
±8kV air
±2kV on AC Mains
±1kV for I/O Lines
±1kV Differential
±2kV Common
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5
Seconds
3A/m
V1 = 3 Vrms
E1 = 3V/m
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
tile. If floors are synthetic, the relative humidity
should be at least 30%.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment.
Mains power quality should be that of a typical
commercial or hospital environment. If the
user of this device requires continued
operation during power mains interruptions, it
is recommended that the device be powered
from an uninterruptible power supply or
battery.
Power frequency magnetic fields should be
that of a typical location in a typical
commercial or hospital environment.
Portable and mobile RF communications
equipment should be separated from the
device by no less than the recommended
separation distances calculated/listed below:
D=(1.2)√ P
D=(1.2)√ P 80 to 800 MHz
D=(2.3)√ P 800 MHz to 2.5 GHz
Where P is the maximum power rating in
watts, and D is the recommended separation
distance in meters.
Field strengths from fixed transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance levels (V1
and E1).
Interference may occur in the vicinity of
equipment marked with the following symbol:
EN - 17

Publicidad

Tabla de contenido

Solución de problemas

loading

Este manual también es adecuado para:

Sleepcube dv56 serieSleepcube dv5hh

Tabla de contenido