Technical Data - SKF TKTL 31 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 32

4. Technical data

Designation
Temperature range
Infrared measurement
Using a probe
Probe supplied
Probe types suitable
Full range accuracy
Tamb = 25°C
Tamb = 23 +/- 3°C
Tamb = 23 +/- 3°C
Tamb = 23 +/- 6°C
Response time
Optics
Distance to spot size
Spectral response
Emissivity variable
Laser
Maximum laser power
Laser pointers
Laser wavelength
Display
Display
Displayed resolution
Controls
Buttons
User selectable backlit display
User selectable laser pointers
(2x)
Measurement modes
Alarm modes
Dimensions
Unit
TL1 Carrying case dimensions
Weight (including case)
6
EN
TKTL 31
-60 to 1 600 °C (-76 to 2 912 °F)
-64 to 1 400 °C (-83.2 to 2 552 °F)
TMDT 2-30, suitable for use up to
900 °C (1 650 °F)
K type probes
Tobj = 15 to 35 °C, +/- 1.0 °C (1.8 °F)
Tobj = 0 to 1 600 °C, +/-2% of reading
or 2 °C (4 °F) whichever is greater
Tobj = -60 to 0 °C, +/- (2 + 0.05/°) °C
With probe: , +/- 1% of reading or
1 °C (1.8 °F) whichever is greater
1 sec
Sense emitted, reflected and transmitted
energy.
75:1
8-14 µm
0.1-1.0
Class 2
1 mW
2 laser dots
635 - 660 nm
LCD black and white display, backlit
0.1 °C/F from -83.2~999.9,
otherwise 1° C/F
4 buttons and 1 trigger.
Function as per instructions for use
yes, On/off
yes, On/off
Max, min, average, differential,
probe/IR dual temperature
High and low level alarm level with
warning bleep
203 × 197 × 47 mm
(8.0 × 7.7 × 1.8 in)
530 × 85 × 180 mm (20.9 × 3.4 × 7.0 in)
1 300 g (2.87 lb)

Publicidad

Tabla de contenido
loading

Tabla de contenido