Apex XT Sense Manual Del Usuario página 23

Ocultar thumbs Ver también para XT Sense:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

XT Sense / English
NOTE 1: U
is the a.c. mains voltage prior to the application of the test level
T
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people
a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters,
an electromagnetic site survey should be considered. If the measured field strength in the location in
which the device is used exceeds the applicable RF compliance level above, the device should be
observed to verify normal operation. If abnormal performance is observed, additional measures may
be necessary, such as reorienting or relocating the device.
b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
Recommended separation distances between portable and mobile RF communications equipment and
this device:
This device is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of this device can help prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment
(transmitters) and this device as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
d
distance
in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
Separation distance according to frequency of transmitter m
150 kHz to 80 MHz
d
=
0.1
0.31
1
3.1
10
80 MHz to 800 MHz
d
=0.6
0.06
0.19
0.6
1.9
6
800 MHz to 2.7 GHz
d
=1.2
0.12
0.38
1.2
3.8
12
20

Publicidad

Tabla de contenido
loading

Tabla de contenido