Initialisation
DTR
RTS
min. 50msec
Setting the baudrate
DTR
RTS
min. 0.5usec
Control commands
The control commands are largely defined by the Standard Commands for
Programmable Instruments (SCPI).
Remarks
ACT-220L/220L+ are programmed by toggling
1
the control lines RTS and DTR. These lines
may not be low at the same time during
0
operation.
In
this
1
220L/220L+ goes in power down mode. If
DTR and RTS are low at the same time or in
0
an undefined condition, both lines must be set
high for at least 50 ms to leave the power
down mode.
Remarks
Before
setting
1
ACT-220L/220L+
according Initialisation. The baudrate is set
0
to 9600 bps according the opposite diagram.
1
A PC usually needs more than 0.5 us for an
I/O-Instruction.
0
47
condition
the
ACT-
the
baudrate
the
has
to
be
initialised