Full-scale value
Set the given pressure value to the normal pressure.
1.
Press the Function key (
2.
the full-scale value will be calibrated and the HM35 returns to normal operation.
Notes
The calibration is always carried out in mbar.
The calibration must take place at a stable room temperature of 22 °C ± 2 °C.
A calibration value will only be accepted if it lies within ± 5 % of the full-scale
value of the HM35.
With the pressure connections open, it is possible to only re-calibrate the zero point.
The date of the last calibration will be stored in the Calibration History.
In case of manual re-calibration, the previous measurement will become invalid.
You should always carry out a complete accuracy check afterwards.
4

Specifications

4.1

Technical data

Measuring media
Media-compatibility
Linearity, hysterisis and
repeatability accuracy
Units
Certification for
Ex/ATEX -Instruments
Operating temperature
Storage temperature
Humidity
Case protection
)
Instrument air or inert gases
all media that is compatible with stainless steel 18/8
(DIN 1.4305)
according to measuring range and use
See Table 4.2
according to measuring range and use
See Table 4.3
EEx ia IIC T4 (resp. T3)
0 °C to 50 °C
-20 °C to 60 °C
max. 95 % rH. (non condensing)
IP 54
54