Setting up the Customised Programs MEM1-3
1. Select MEM1 for a constant program by pressing the PRG button on
the front panel.
2. Hold down the SET button for three seconds to enter the setup menu.
The Hz symbol will flash.
3. Press the A+ or A- button select the desired Rate (Hz) from 2 to 100 Hz.
4. Press the SET button to select the Pulse width (µS). The µS symbol
will flash.
5. Press the A+ or A- button to select the desired Pulse Width from
50 to 450 µS.
6. Press the SET button to select the time. The MIN symbol will flash
and number of minutes will appear on the display.
7.
Press the A+ or A- button to select the desired time from 5 to 60 minutes.
8. Press the SET button to select the work time in seconds WORK will
appear on the display. SEC symbol will flash.
9. Press the A+ or A- button to select the desired work time from
2 to 99 seconds.
10. Press the SET button to select the rest time in seconds REST will
appear on the display. SEC symbol will flash
11. Press the A+ or A- button to select the desired rest time from
2 - 99 seconds.
12. Press the SET button to select the ramp up time in seconds RAMP
will appear on the display. SEC symbol will flash
13. Press the A+ or A- button to select the desired ramp time from
0.2 - 9.9 seconds.
14. Pressing the SET button again will bring you back to the Rate setting.
15. Press the PRG button to save the program. Repeat the above
procedure to re-program.
Note: You must press the PRG button before locking the unit.