Table 201: Guidance and manufacturer's declaration
– electromagnetic emissions
The PERICALM™ product is intended for use in the electromagnetic
environment specified below. The customer or the user of the PERICALM™
product
should ensure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Table 202: Guidance and manufacturers declaration
– electromagnetic immunity
PERICALM™
T he
product is intended for use in t he electromagnetic environment specified
below. T he customer or t he user of the
such an environment , and that precautions regarding t hat environment are heeded.
Immunity test
IEC 60601
test level
Electrostat ic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
PERICALM™ Instruction Manual - GB
Compliance
Electromagnetic environment –
Group 1
T he PERICALM™ product uses RF
energy only for its internal function.
T herefore, its RF emissions are very
low and are not likely to cause any
interference in nearby electronic
equipment.
T he PERICALM™ product is suitable
Class B
for use in all establishments,
including domestic establishments
Not applicable
and those directly connected to the
public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Not applicable
PERICALM™
product should ensure that it is used in
Compliance
Electromagnetic environment –
le vel
±6 kV cont act
Floors should be wood, concrete or
±8 kV air
ceramic t ile. If floors are covered with
synthetic material, the relative
humidity should be at least 30 %.
3 A/m
Power frequency magnetic fields should
be at characteristic levels of a typical
locat ion in a t ypical commercial or
hospital environment.