Note Ensure RTAs are set to zero before performing a calibration.
• To perform a calibration, press
Internal RTD
External RTD
• Press
on the left of the desired temperature sensor to display:
Note For ease of calibration, ensure the sensor selected is the sensor controlling the
temperature. Do not calibrate the internal probe if the external probe is controlling
temperature.
on the right of Calibrate to display:
• Press
• Press
on the right of SPx to display the Change of Value Display. Enter
the desired high of low calibration temperature. Press
on the right of Calibration to display:
Calibration
Internal RTD
Restore User Cal
Save User Cal
Restore Factory Cal
Internal RTD
Section 7 Password
Calibrate
Internal RTD
xx.xx°C
High
xx.xx°C
Low
x.xx°C
Calibrate
SP x
xx.xx °C
.
7-3