CAUTION
1. Never use the same SENSORCAP on more than one patient. Attempting to wipe the SENSORCAP may damage
it and may cause instrument malfunction.
2. It is usually considered safe to use the same SENSORCAP on both eyes of the same patient This may help expe-
diting your work with the patient and help avoid interruption in your patient session.
3. Using the SENSORTIPS without SENSORCAPS will cause them to deteriorate quickly, and may necessitate
expensive replacement. Working with unprotected SENSORTIPS exposes patient to risk of cross-contamination
from previous patients.
4. Never disinfect SENSORCAPS with alcohol or other disinfectants as this dissolves the silicone. Sterility and
impermeability cannot be guaranteed in this case.
Icons on labels
Disposal
Used SENSORCAPS tip covers should be disposed of with other medical office waste. All disposals should be carried
out in accordance with applicable local and national regulations.
Packaging and ordering
One box of SENSORCAPS contains 240 sterile disposable SENSORCAPS in APPLIPACK packaging (Part number
311.701.201). May be ordered from your local distributor or from Ziemer Ophthalmic Systems AG, Allmendstrasse
11, 2562 Port, Switzerland, Email: support@ziemer-ophthalmics.com, Fax: +41 (0)32 332 70 71.
Disclaimer of liability
SMT AG shall not be liable for any inquiry or damage suffered by a patient as a result of any application technique
used by a physician or qualified staff or any prescription or selection and use for this product for a particular patient.
The physician or qualified staff takes full responsibility for the proper use of this product and the technique used
for each patient.
Warranty information
Please consult the Limited Warranty Terms, which are included in the PASCAL® Dynamic Contour Tonometer User
Manual.
DFU_610-50-0002_PASCAL.indd 2
When placing the SENSORTIP back into its storage container, make sure to take off the SENSORCAP.
Leaving a SENSORCAP on the SENSORTIP may cause damage to the SENSORTIP when trying to
remove the SENSORCAP after storage. The SENSORCAP may tear, leaving difficult to remove
residues on the surface of the SENSORTIP.
Single use
All items in the Procedure Pack are single use and must be discarded after use.
Attention or read instructions
Warning! Important information. Read instructions.
Manufacturer
Name and address of the manufacturer
Use by date /Expiration date
Indicates the date after which the medical device is not be used
Batch Code
Indicates the manufacturers batch code so that the batch can be identified
Sterilized medical device
The medical device has been sterilized with irradiation
Catalog number
Indicates the manufacturers catalog number