− Remove all the packaging and check that the appliance is in perfect conditions. In case
of visible damage, do not connect the appliance and notify the sales point immediately.
− Remove the PVC protection from the panels.
− Dispose of packaging according to regulations. Generally material is divided according
to composition and should be delivered to the waste disposal service.
− There are no special instructions regarding distances from other appliances or walls,
however it is advisable to maintain a sufficient distance to allow any servicing
operations to be performed. In the event the appliance should be installed in direct
contact with inflammable walls, it is advisable to fit a suitable heat insulation.
− The flooring must be particularly compact and resistant in order to withstand the
dynamic load of the apparatus while it is working. The floor slab underneath the
apparatus must be at least in cement B25.
Warning! The appliance must stand level. A significantly uneven or sloping stance
can affect the operation of the appliance adversely.
Warning! The apparatus is equipped with a control board for wall fitting which must
be easy to reach for the operator.
Warning! The apparatus must be fixed on the floor by means of three expansion
bolts M8x150 as shown in the illustration (supplied together with machine).
In order to avoid water infiltrations inside the basement, seal the gasket on the base
of the machine with silicone glue before fixing the three expansion bolts.
Instruction manual for installation, maintenance and use
Centrifuge DRC 700V
2.1. POSTIONING
DRC700V-GB-00
Page