GUIDANCE AND MANUFACTURER'S DECLARATION – ELECTROMAGNETIC IMMUNITY
GUÍA Y DECLARACIÓN DE ORTOTECSA – INMUNIDAD ELECTROMAGNÉTICA
The paraffin bath is intended for use in an electromagnetic environment specified below. The customer or user of the
paraffin bath should assure that it is used in such an environment
Immunity
IEC 60601 test
Test
Level
Ensayo de
Nivel de ensayo
inmunidad
de la norma
Conducted RF
3 v
rms
IEC 61000-4-6
150 kHz to 80 MHz
RF conducida
3 v/m
Radiated RF
80 MHz to 2,5 GHz
IEC 61000-4-3
RF radiada
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To access the
Ortotecsa,s.l. C/Peña de Francia,8
Tlf.:+34918719132
Compliance
Level
Nivel de
conformidad
3 V
rms
3 V/m 80 MHz to 1
GHz (due to the
testing laboratory
limitations)
(debido a las
limitaciones de
laboratorio)
28500 Arganda del Rey (Madrid), España.
Fax: +34918719323
email:
INSTRUCCIONES DE USO Y MANTENIMIENTO
USER AND MAINTENANCE MANUAL
Last Revision 12/03/2020
Electromagnetic Environment
Guidance
Entorno electromagnético, guía
Portable and mobile RF communications
equipment should be used no closer to any part
of the infrared lamp, including cables, tan the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter.
Los equipos móviles y comunicaciones de RF
no se deben usar más cerca de cualquier parte
de la lámpara, incluyendo los cables, que la
distancia de separación recomendada a la
frecuencia del transmisor.
Recommended separation distance:
Distancia de separación recomendada:
d = 1,17 √P 150 kHz to 80 MHz
d =1,17√P 80 MHz to 800 MHz
d=2,33√P 800 MHz to 2,5 GHz
Where P is the maximum output power rating
of the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Donde p es la maxima potencia de salida
asignada de transmisor en vatios (w)
conforme al fabricante del transmisor y d
es la distancia de separación recomendada
en metros.
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey a,
should be less than the compliance level in
each frequency range.
Las intensidades de campo desde el transmisor
fijo de RF, según se determina por un estudio
electromagnético de lugar, debe ser menor que
el nivel de conformidad en cada rango de
frecuencia.
Interference may occur in the vicinity of the
equipment marked with the following symbol:
La interferencia puede ocurrir en la vecindad
del equipo marcado con el siguiente símbolo:
info@ortotecsa.com
www.ortotecsa.com
b
Página 15 de 37