Connection cable

Erbe
Length
connecting cable
20192-101
4 m
20192-112
4 m
6

How to use

6.1

Only use sterile products

Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
6.2

Check the product

Before each use, check the product including all the insulation
for signs of physical damage!
Check the function of the needle before each use. It must be pos-
sible to fully extend the needle distally by pressing the handle. If
the pressure is set, the needle must fully retract back into the in-
strument sheath. The needle may no be bent. The insulation of
the needle must be intact.
If damaged, do not use this product!